Title (with LTR): Precision Xceed Glucose Procedure (LTR19678)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:20:23 PM)
6
6. Touch the blood drop to white target area at tip
of test strip while the “Apply Sample” prompt
appears in the display window.
Hold finger in place until test begins.
Never add additional sample. Discard strip
and use a new one if needed.
7. If necessary, the blood can be collected in a
heparin-coated capillary tube and then applied
to the test strip within 30 minutes of collection.
Collecting Arterial Blood Samples
With meter ready, obtain a sample:
Item Comment(s)
Clear the arterial line before drawing a blood
sample. Withdraw blood sample into a syringe
that contains sodium or lithium heparin.
Caution: You could obtain a false low glucose
result if:
1. sufficient flush is not withdrawn from the
arterial line (actual volume required to be
removed is dependent upon length of
line), or
2. if liquid heparin is used to anticoagulate
the sample and the ratio of blood to
heparin is not controlled or not taken into
consideration.
Use the sample within 30 minutes of collection.
Mix the syringe several times immediately
before applying the sample to the target area
on the test strip.
Allow a drop of blood to form at the tip of the
syringe